6½ 位精度,5.3 位下 15,000 读数/秒,10.9 英寸触摸屏趋势图,台式测量效率之选。
| Range | Resolution | Input Impedance | 24 Hours T ±1°C CAL | 90 Days T ±5°C CAL | 1 Year T ±5°C CAL | 2 Years T ±5°C CAL | Temperature Coefficient |
| 100 mV | 100 nV | >10 GΩ or 10 MΩ ±1% | 0.0015 + 0.0030 | 0.0025 + 0.0035 | 0.0030 + 0.0035 | 0.0035 + 0.0035 | 0.0001 + 0.0005 |
| 1 V | 1 μV | >10 GΩ or 10 MΩ ±1% | 0.0015 + 0.0006 | 0.0020 + 0.0006 | 0.0025 + 0.0006 | 0.0030 + 0.0006 | 0.0001 + 0.0001 |
| 10 V | 10 μV | >10 GΩ or 10 MΩ ±1% | 0.0010 + 0.0004 | 0.0020 + 0.0005 | 0.0025 + 0.0005 | 0.0030 + 0.0005 | 0.0001 + 0.0001 |
| 100 V | 100 μV | 10 MΩ ±1% | 0.0015 + 0.0006 | 0.0035 + 0.0006 | 0.0040 + 0.0006 | 0.0050 + 0.0006 | 0.0006 + 0.0001 |
| 1000 V 1 | 1 mV | 10 MΩ ±1% | 0.0020 + 0.0006 | 0.0035 + 0.0006 | 0.0040 + 0.0006 | 0.0050 + 0.0006 | 0.0006 + 0.0001 |
| Measurement Rate in NPLCs | Digits | DCV RMS Noise Uncertainty (in % of range + fixed base) 2 | NMRR 3 | CMRR 3 |
| 5 4 | 6.5 | 0 | 100 dB | 140 dB |
| 1 4 | 0 | 90 dB | 140 dB |
| 0.1 4 | 0.00015 + 1 µV | 40 dB | 120 dB |
| 0.1 | 5.5 | 0.00015 + 4 µV | -- | 120 dB |
| 0.01 | 0.00030 + 6 µV | -- | 80 dB |
| 0.0005 | 4.5 | 0.00500 + 40 µV | -- | 80 dB |
| Scanner Card | Add the Following Uncertainty | Maximum Input Signal Level |
| 2000-SCAN | 1 μV | 110 V |
| 2001-TCSCAN | 1 μV | 110 V |
| Range | Resolution | Test Current (±5%) | Open Circuit Voltage (±5%) | 24 Hours T ±1°C CAL | 90 Days T ±5°C CAL | 1 Year T ±5°C CAL | 2 Years T ±5°C CAL | Temperature C |
| 1 Ω 6 | 1 μΩ | 10 mA | 12.5 V | 0.0080 + 0.0200 | 0.0080 + 0.0200 | 0.0085 + 0.0200 | 0.0100 + 0.0200 | 0.0006 + 0.0010 |
| 10 Ω 6 | 10 μΩ | 10 mA | 12.5 V | 0.0020 + 0.0020 | 0.0080 + 0.0020 | 0.0085 + 0.0020 | 0.0100 + 0.0020 | 0.0006 + 0.0001 |
| 100 Ω | 100 μΩ | 1 mA | 9.2 V | 0.0020 + 0.0020 | 0.0075 + 0.0020 | 0.0085 + 0.0020 | 0.0100 + 0.0020 | 0.0006 + 0.0001 |
| 1 kΩ | 1 mΩ | 1 mA | 9.2 V | 0.0020 + 0.0006 | 0.0065 + 0.0006 | 0.0075 + 0.0006 | 0.0090 + 0.0006 | 0.0006 + 0.0001 |
| 10 kΩ | 10 mΩ | 100 μA | 12.7 V | 0.0020 + 0.0006 | 0.0065 + 0.0006 | 0.0075 + 0.0006 | 0.0090 + 0.0006 | 0.0006 + 0.0001 |
| 100 kΩ | 100 mΩ | 10 μA | 12.5 V | 0.0020 + 0.0006 | 0.0070 + 0.0010 | 0.0075 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
| 1 MΩ | 1 Ω | 10 μA | 12.5 V | 0.0020 + 0.0006 | 0.0075 + 0.0006 | 0.0100 + 0.0006 | 0.0120 + 0.0006 | 0.0006 + 0.0001 |
| 10 MΩ 7 | 10 Ω | 0.7 μA || 10 MΩ | 7.1 V | 0.0150 + 0.0006 | 0.0200 + 0.0010 | 0.0400 + 0.0010 | 0.0450 + 0.0010 | 0.0105 + 0.0001 |
| 100 MΩ 7 | 100 Ω | 0.7 μA || 10 MΩ | 7.1 V | 0.0800 + 0.0030 | 0.2000 + 0.0030 | 0.2000 + 0.0030 | 0.2500 + 0.0030 | 0.047 + 0.0001 |
| Measurement Rate in NPLC | Digits | 2-wire RMS Noise Uncertainty (in % of range + fixed base) | 4-wire RMS Noise Uncertainty, Offset Compensation OFF (in % of range + fixed base) 9 | 4-wi |
| 0.1 10 | 0.00015 + 0.10 mΩ | 0.00020 + 0.20 mΩ | 0.00030 + 0.25 mΩ |
| 0.1 | 5.5 | 0.00050 + 0.35 mΩ | 0.00180 + 2.00 mΩ | 0.00350 + 3.50 mΩ |
| 0.01 | 0.00070 + 0.50 mΩ | 0.00260 + 2.50 mΩ | 0.00500 + 4.00 mΩ |
| 0.0005 | 4.5 | 0.00650 + 3.50 mΩ | 0.01000 + 7.00 mΩ | 0.01500 + 10.00 mΩ |
| Scanner Card | Contact Resistance |
| 2000-SCAN | 1 Ω at end of life |
| 2001-TCSCAN | 1 Ω at end of life |
| Range | Resolution | Burden Voltage | 24 Hours T ±1°C CAL | 90 Days T ±5°C CAL | 1 Year T L ±5°C CA | 2 Years T ±5°C CAL | Temperature Coefficient |
| 10 μA | 10 pA | <0.13 V | 0.007 + 0.002 | 0.035 + 0.005 | 0.045 + 0.005 | 0.055 + 0.005 | 0.0030 + 0.0006 |
| 100 μA | 100 pA | <0.14 V | 0.010 + 0.002 | 0.035 + 0.005 | 0.045 + 0.005 | 0.055 + 0.005 | 0.0020 + 0.0005 |
| 1 mA | 1 nA | <0.17 V | 0.007 + 0.006 | 0.035 + 0.005 | 0.045 + 0.005 | 0.055 + 0.005 | 0.0020 + 0.0005 |
| 10 mA | 10 nA | <0.17 V | 0.006 + 0.003 | 0.018 + 0.005 | 0.020 + 0.005 | 0.025 + 0.005 | 0.0015 + 0.0005 |
| 100 mA | 100 nA | <0.20 V 11 | 0.010 + 0.003 | 0.015 + 0.005 | 0.020 + 0.005 | 0.025 + 0.005 | 0.0015 + 0.0005 |
| 1 A | 1 μA | <0.55 V 11 | 0.020 + 0.004 | 0.030 + 0.005 | 0.040 + 0.005 | 0.050 + 0.005 | 0.0030 + 0.0005 |
| 3 A | 1 μA | <1.70 V 11 | 0.030 + 0.004 | 0.040 + 0.004 | 0.050 + 0.004 | 0.060 + 0.004 | 0.0030 + 0.0005 |
| 10 A 12 | 10 μA | <0.50 V | 0.140 + 0.025 | 0.190 + 0.025 | 0.220 + 0.025 | 0.250 + 0.025 | 0.0060 + 0.0005 |
| 10 μA | 100 μA | 1 mA | 10 mA | 100 mA | 1 A | 3 A | 10 A |
| 10 kΩ | 1 kΩ | 100 Ω | 10 Ω | 1 Ω | 100 mΩ | 100 mΩ | 5 mΩ |
| Measurement Rate in NPLC | Digits | DCI RMS Noise Uncertainty (in % of range + fixed base) |
| 0.1 15 | 0.0009 + 10.0 pA |
| 0.1 | 5.5 | 0.0015 + 5.0 nA |
| 0.01 | 0.0030 + 5.0 nA |
| 0.0005 | 4.5 | 0.0200 + 5.0 nA |
| Type | Resolution | Range | 2 Year Accuracy T ±5°C; all uncertainties in °C CAL | Temperature Coefficient in°C/°C |
| Measurement Method | Resolution | Range | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient in°C/°C |
| 2-wire 17 | 0.01°C | –200° to 850°C | 0.80 | 0.003 |
| 3-wire 18 | 0.01°C | –200° to 600°C | 0.35 | 0.003 |
| 4–wire | 0.01°C | –200° to 600°C | 0.06 | 0.003 |
| Measurement Method | Resolution | Range | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient in°C/°C |
| 2-wire | 0.01°C | –80° to 150°C | 0.08 | 0.002 |
| Thermistor Type | Common Model Number | 70° to 100°C | >100° to 150°C |
| 2.2 kΩ | 44004 | 0.22°C per Ω | 1.11°C per Ω |
| 5 kΩ | 44007 | 0.10°C per Ω | 0.46°C per Ω |
| 10 kΩ | 44006 | 0.04°C per Ω | 0.19°C per Ω |
| Range | Resolution | Calibration Cycle | 3 Hz to 5 Hz | 5 Hz to 10 Hz | 10 Hz to 20 kHz | 20 kHz to 50 kHz | 50 kHz to 100 kHz | 100 kHz to 300 kHz |
| 100 mV | 100 nV 1 µV 10 µV 100 µV 100 µV | 24 hours | 1.00 + 0.02 | 0.35 + 0.02 | 0.04 + 0.02 | 0.10 + 0.04 | 0.55 + 0.08 | 4.00 + 0.50 |
| Temperature Coefficient | 0.100 + 0.003 | 0.035 + 0.003 | 0.005 + 0.003 | 0.011 + 0.005 | 0.060 + 0.08 | 0.200 + 0.020 |
| Range | Resolution | Burden Voltage | Frequency | 24 Hours T ±1°C CAL | 90 Days T ±5°C CAL | 1 Year T ±5°C CAL | 2 Years T ±5°C CAL | Temperature Coefficient |
| 100 µA | 100 pA | <0.14 V | 3 Hz – 1 kHz | 0.10 + 0.07 | 0.10 + 0.07 | 0.10 + 0.07 | 0.10 + 0.07 | 0.015 + 0.010 |
| 1 mA | 1 nA | <0.17 V | 3 Hz – 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 |
| 10 mA | 10 nA | <0.17 V | 3 Hz – 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.10+ 0.04 | 0.015 + 0.006 |
| 100 mA | 100 nA | <0.20 V 23 | 3 Hz – 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 |
| 1 A | 1 µA | <0.75 V 23 | 3 Hz – 5 kHz 24 | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 |
| 3 A | 1 µA | <1.70 V 23 | 3 Hz – 5 kHz 24 | 0.15 + 0.06 | 0.15 + 0.06 | 0.15 + 0.06 | 0.15 + 0.06 | 0.015 + 0.006 |
| 10 A | 10 µA | <0.50 V | 3 Hz – 1 kHz 24 | 0.40 + 0.06 | 0.40 + 0.06 | 0.40 + 0.06 | 0.40 + 0.06 | 0.015 + 0.006 |
| Range | Resolution | Frequency | Period | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient in°C/°C |
| 100 mV to 750 V (For signals >5% of range and >10 mV ) RMS | 0.0001% of reading | 3 Hz to 10 Hz | 333 ms to 100 ms | 0.100 | 0.0002 |
| Range | Resolution | Test Current | Open Circuit Voltage (±5%) | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient |
| 1 kΩ | 100 mΩ | 1 mA | 9.2 V | 0.010 + 0.010 | 0.0006 + 0.0001 |
| Range | Resolution | Charge Current (±5%) 31 | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient |
| 1 nF | 0.1 pF | 1 µA | 0.80 + 0.50 | 0.05 + 0.05 |
| 10 nF | 1 pF | 10 µA | 0.40 + 0.10 | 0.05 + 0.01 |
| 100 nF | 10 pF | 100 µA | 0.40 + 0.10 | 0.05 + 0.01 |
| 1 µF | 0.1 nF | 100 µA | 0.40 + 0.10 | 0.05 + 0.01 |
| 10 µF | 1 nF | 1 mA | 0.40 + 0.10 | 0.05 + 0.01 |
| 100 µF | 10 nF | 1 mA | 0.40 + 0.10 | 0.05 + 0.01 |
| Voltage Measure Range | Resolution | Maximum Voltage Measurement | Test Current (±5%) | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient |
| 10 V | 10 μV | 12 V | 10 µA | 0.0045 + 60.0 µV | 0.0008 + 10 µV |
| Range | Resolution | Input Impedance | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient |
| 100 mV | 10 µV | >10 GΩ or 10 MΩ ±1% | 0.040 + 0.020 | 0.0025 + 0.0030 |
| 1 V | 100 µV | >10 GΩ or 10 MΩ ±1% | 0.030 + 0.010 | 0.0025 + 0.0010 |
| 10 V | 1 mV | >10 GΩ or 10 MΩ ±1% | 0.030 + 0.010 | 0.0025 + 0.0010 |
| 100 V | 10 mV | 10 MΩ ±1% | 0.030 + 0.010 | 0.0025 + 0.0010 |
| 1000 V | 100 mV | 10 MΩ ±1% | 0.030 + 0.010 | 0.0025 + 0.0010 |
| Range | Resolution | Burden Voltage | 2 Year Accuracy T ±5°C CAL | Temperature Coefficient |
| 100 µA | 10 nA | <0.14 V | 0.07 + 0.05 | 0.0030 + 0.0035 |
| 1 mA | 100 nA | <0.17 V | 0.07 + 0.03 | 0.0030 + 0.0035 |
| 10 mA | 1 µA | <0.17 V | 0.05 + 0.03 | 0.0030 + 0.0035 |
| 100 mA | 10 µA | <0.20 V 34 | 0.05 + 0.03 | 0.0020 + 0.0035 |
| 1 A | 100 µA | <0.55 V 34 | 0.07 + 0.03 | 0.0040 + 0.0035 |
| 3 A | 100 µA | <1.70 V 34 | 0.09 + 0.04 | 0.0040 + 0.0035 |
| 10 A | 1 mA | <0.50 V | 0.25 + 0.08 | 0.0060 + 0.0100 |
| Function: Range | Spur-free Range SFDR (1 kHz / 10 kHz / 50 kHz) | THD + Noise SNDR (1 kHz / 10 kHz / 50 kHz) | Bandwidth (–3 dB, 5%) | Effective Number of Bits (1 kHz/10 kHz/50 |
| DCV: 100 mV | 75 / 70 / 50 | 65 / 60 / 50 | 210 kHz | 9 / 9 / 7 |
| DCV: 1 V | 95 / 90 / 75 | 80 / 80 / 75 | 210 kHz | 12 / 12 / 11 |
| DCV: 10 V | 95 / 80 / 70 | 90 / 80 / 70 | 440 kHz | 13 / 12 / 10 |
| DCV: 100 V | 50 / 35 / 25 | 50 / 40 / 30 | 17 kHz | 10 / 8 / 7 |
| DCV: 1000 V | 50 / 35 / 25 | 50 / 40 / 30 | 17 kHz | 13 / 11 / 10 |
| DCI: 100 µA | 80 / 65 / 45 | 70 / 65 / 45 | 430 kHz | 12 / 10 / 8 |
| DCI: 1 mA | 80 / 65 / 45 | 70 / 65 / 45 | 570 kHz | 12 / 10 / 8 |
| DCI: 10 mA | 80 / 65 / 45 | 70 / 65 / 45 | 230 kHz | 12 / 10 / 8 |
| DCI: 100 mA | 80 / 65 / 45 | 70 / 65 / 45 | 340 kHz | 12 / 10 / 8 |
| DCI: 1 A | 70 / 50 / 40 | 65 / 50 / 40 | 25 kHz | 11 / 8 / 7 |
| DCI: 3 A | 70 / 50 / 40 | 65 / 50 / 40 | 25 kHz | 11 / 8 / 7 |
| DCI: 10 A | 45 / 25 / 20 | 43 / 30 / 30 | 40 kHz | 7 / 5 / 5 |
| Method | Measurement |
| Channel Ratio (through rear input scanner card) | _C_h__a_n_n_e__l _A_ C hannel Ratio = Channel B Accuracy = (Accuracy of channel A measure range + Accuracy of channel B measure range) × Channel ratio |
| Channel Average (through rear input scanner card) | _C_h__a_n_n_e__l _A_ +__ C_h__a_n_n_e__l _B_ C hannel Average = 2 Accuracy = Accuracy of channel A measure range + Accuracy of paired channel B measure range |
| DCV Input Ratio (HI-LO/SHI–SLO) 36 | Ratio = ________H_I_ s_i_g_n__a_l _______ SHI signal – SLO signal Accuracy = ( _H_I_ r_a_n__g_e_ × DCV% of range accuracy + __________1_0_ _V_ _________ × 0.000 |
| NPLC | Functions: DCV (10 V) 2-wire Ω (≤10 kΩ), DCI (1 mA) | Functions: 4-wire Ω (≤1 kΩ) 4-wire and 3-wire RTD | Function: Thermistor or Thermocouple |
| 0.1 | 584 (490) | 440 (380) | 180 (160) | 170 (150) | 580 (480) | 440 (380) |
| 0.01 | 4900 (4100) | 4800 (4100) | 400 (390) | 400 (390) | 4800 (4100) | 4700 (4000) |
| 0.0005 | 20600 (20600) | 19800 (19800) | 460 (460) | 460 (460) | 21000 (21000) | 20300 (20300) |
| Function: ACV, ACI | Function: Frequency, Period | Measurements (readings per second) |
| Detector Bandwidth | Aperture | Buffer or Computer |
| 3 Hz | 200 ms | 1 |
| 30 Hz | 20 ms | 10 |
| 300 Hz | 2 ms | 100 |
| Typical Scanning Measurement Rates | Measurements Into Buffer/Computer (channel per second) |
| Scanning DCV or 2-wire Ω | >90 with 2000-SCAN card, >90 with 2001-TCSCAN card |
| Scanning Thermocouple, Thermistor, or 2-wire RTD | >85 with 2000-SCAN card, >85 with 2001-TCSCAN card |
| Scanning 4-wire Ω and 3- or 4-wire RTD | >80 with 2000-SCAN card, >80 with 2001-TCSCAN card |
| Scanning ACV | >60 with 2000-SCAN card, >60 with 2001-TCSCAN card |
| Scanning Alternating DCV and 2-wire Ω | >85 with 2000-SCAN card, >85 with 2001-TCSCAN card |
| Function | Function Change Time 42 | Range Change Time 43 | Autorange Time 42 |
| DCV, DCI, or 2-wire Ω 44 | <4 ms | <1.3 ms | <3.2 ms |
| 4-wire Ω 45 or 3-wire RTD | <5.5 ms |
| Thermistor | — |
| Frequency or Period (2 ms aperture) | <1800 ms | <50 ms 46 | <50 ms 46 |
| ACI (300 Hz bandwidth) | <100 ms | <4 ms | <5 ms |
| Capacitance | <4 ms | <3 ms | <30 ms |
| Digitize | <4 ms | <5 ms | — |
| Diode | <11 ms | — | — |
| Continuity | <11 ms | — | — |
| Thermocouple | <4 ms | — | — |
| Average for 1000 readings (binary) | 441,000 | 268,000 | 201,000 | 10,000 |
| Average for 1000 readings with relative timestamp (binary) | 272,000 | 150,000 | 105,000 | 2,900 |
| Average for 1000 readings with formatted elements 48 | 46,000 | 29,000 | 17,000 | 290 |
| Sampling rate | Measurements over USB to computer (readings per second) |
| 10 kS/s | Up to 10,000 |
| 50 kS/s | Up to 50,000 |
| 100 kS/s | Up to 100,000 |
| 1 MS/s up to 7 s maximum duration | At least 90,000 |
问:DMM6500的分辨率达到什么水平?
答:DMM6500 的分辨率为 6½ 位图形触摸屏台式/系统万用表。以上为官方规格,实际供货配置与选型可让工程师按你的测试需求核对。
问:DMM6500的测量功能达到什么水平?
答:DMM6500 的测量功能为 DCV / ACV / DCI / ACI / 2Ω 与 4Ω 电阻 / 电容 / 温度 / 频率 / 周期 / 二极管测试 / 数字化采集。以上为官方规格,实际供货配置与选型可让工程师按你的测试需求核对。
问:DMM6500的触摸屏达到什么水平?
答:DMM6500 的触摸屏为 5 英寸电容触控。以上为官方规格,实际供货配置与选型可让工程师按你的测试需求核对。
问:DMM6500主要用在什么测量场合?
答:6½ 位精度,5.3 位下 15,000 读数/秒,10.9 英寸触摸屏趋势图,台式测量效率之选。具体是否匹配你的测试对象,可把需求发给工程师免费核对型号。
问:DMM6500和同系列其他型号怎么选?
答:建议先看 分辨率(6½ 位图形触摸屏台式/系统万用表)、测量功能(DCV / ACV / DCI / ACI / 2Ω 与 4Ω 电阻 / 电容 / 温度 / 频率 / 周期 / 二极管测试 / 数字化采集) 这几项,再结合你的被测信号/器件类型确认档位。不同子型号配置有差异,选型时让工程师按需求比对最稳妥。
问:DMM6500 是全新原装吗?货期和保修怎么算?
答:全新原装正品、正规渠道供货,序列号可送原厂查验(非翻新、非演示机);常规配置现货或 1~2 周内发货,特殊配置下单前书面确认交期;按厂家政策享受原厂保修。